* Characteristics of auditory ALARM SIGNALS - Rational
ü
An ALARM SYSTEM provided
with auditory ALARM SIGNALS shall have at least one set of ALARM SIGNALS that either:
a) is priority encoded and meets the requirements of Table 3 and Table 4; (Rational), or
b) is generated by means of different technology (e.g., voice synthesizing of verbal ALARM SIGNALS) and is VALIDATED (e.g., by clinical or simulated clinical USABILITY testing) (Rational).
ü
If the ALARM SYSTEM is
additionally provided with other sets of auditory ALARM SIGNALS, the following
shall apply (Rational):
c) auditory ALARM SIGNALS shall be priority encoded;
d) HIGH PRIORITY auditory ALARM SIGNALS of a particular set of ALARM SIGNALS shall convey a higher level of urgency than the MEDIUM or LOW PRIORITY ALARM SIGNALS and INFORMATION SIGNALS of that ALARM SIGNAL set;
e) MEDIUM PRIORITY auditory ALARM SIGNALS of a particular set of ALARM SIGNALS shall convey a higher level of urgency than the LOW PRIORITY ALARM SIGNALS and INFORMATION SIGNALS of that ALARM SIGNAL set;
f) auditory ALARM SIGNALS shall be VALIDATED, e.g., by clinical or simulated clinical USABILITY testing;
g) means shall be provided to store a set of auditory ALARM SIGNALS in the DEFAULT ALARM PRESET (Rational); and
h) means may be provided to store a set of auditory ALARM SIGNALS in any ALARM PRESET (Rational).
ü
NOTE 1 See also Annex D.
ü
NOTE 2 Attention is
drawn to IEC 60601-1-6.
ü
Any melody shall
preclude the possibility of confusion with the auditory ALARM SIGNALS of Table 3, Table 4 and Annex F, unless their meaning is the same.
? If
any of the melodies of Annex F is used to meet the requirements of Table 3, Table 4, its meanings
shall be as specified in Annex F.
בעיית ניסוח,
או אי הבנה
שלי?
ü When a TECHNICAL ALARM CONDITION that precludes the generation of the usual ALARM SIGNALS occurs, e.g. power or ALARM SYSTEM failure, the ALARM SYSTEM may generate an auditory ALARM SIGNAL that does not comply with the above requirements.
ü
If selection of auditory
ALARM SIGNAL sets is provided, means shall be provided for the RESPONSIBLE
ORGANIZATION to prevent the OPERATOR from unauthorized access to changing the
auditory ALARM SIGNAL set in use (see 6.7).
היצרן צריך לספק מנגנון הגנה, למשל, על ידי סיסמא
ü
Compliance
is checked by inspection and functional testing of the ALARM SYSTEM and
inspection of any relevant VALIDATION documentation.
אפשר לדרוש
מהארגון
לחזור על
בדיקות
השימושיות של
היצרן. היצרן יכול
לספק ללקוח סט
של תרחישים
ותוכנת הדמיה,
על מנת שילמד
להכיר את
ההתרעות מראש.