6.1.1 * General - Rational יש שאלות
ü
If ALARM CONDITIONS are grouped into PHYSIOLOGICAL
ALARM CONDITIONS, TECHNICAL ALARM
CONDITIONS or other ALARM CONDITION groups by the MANUFACTURER, this shall be disclosed in the
instructions for use.
ü
Compliance
is checked by inspection of the instructions for use.
6.1.2
* ALARM CONDITION priority - Rational
ü
ALARM CONDITIONS shall be assigned to one or more of the
following priorities: HIGH PRIORITY, MEDIUM PRIORITY, or LOW PRIORITY. Unless a particular ALARM CONDITION priority is specified in a relevant
particular standard, the assignment of priorities is part of the RISK MANAGEMENT
PROCESS and shall be
based on Table 1. The priority of each ALARM CONDITION shall be disclosed in the instructions
for use. Priorities may be identified in groups.
ü
Compliance
is checked by inspection of the instructions for use and RISK MANAGEMENT FILE.
Table 1 – ALARM CONDITION
priorities