Subclause
6.1.2 – ALARM CONDITION priority
ü
ALARM CONDITIONS should be prioritized based on the urgency of the required OPERATOR response or awareness of
the situation that triggered the ALARM
CONDITION. Priority is assigned
through RISK ANALYSIS, either by the writer(s) of a particular standard or by the MANUFACTURER.
ü
NOTE Some ALARM
SYSTEMS have OPERATOR-configured or RESPONSIBLE ORGANIZATION-configured
priorities.
ü
MANUFACTURERS assign ALARM CONDITION priorities
based on RISK ANALYSIS. This RISK ANALYSIS should
primarily consider the severity and rapidity of onset of HARM if the ALARM
CONDITION is not corrected. It should also consider other
factors such as the sensitivity and specificity of the ALARM CONDITION for the
actual event in the PATIENT or the equipment. The level of the priority of ALARM SIGNAL only
suggests to the OPERATOR the speed at which the OPERATOR should respond to, or be aware of, an ALARM CONDITION. The
actual speed of response or awareness required is ultimately based on the
assessment by the OPERATOR.
ü
“Immediate”
category problems are those that are likely to cause PATIENT injury or
death within seconds to several minutes if uncorrected. Few problems fall into
the “immediate” category.
EXAMPLE 1 Asystole
EXAMPLE 2 Ventricular fibrillation
EXAMPLE 3 Failure of a cardiac support device (intra-aortic balloon
pump, cardiopulmonary bypass machine)
EXAMPLE 4 Sustained high airway pressure
EXAMPLE 5 Extreme hypoxemia
EXAMPLE 6 Sustained high-energy radiation beam
ü
“Prompt”
category problems, on the other hand, do not cause PATIENT injury or
death until at least several to many minutes have elapsed.
EXAMPLE 7 Many cardiac arrhythmias
NOTE Most cardiac arrhythmias would be prompt or delayed.
EXAMPLE 8 High or low blood pressure
EXAMPLE 9 Apnea (unless prolonged or associated with extreme hypoxia)
EXAMPLE 10 Mild hypoxemia
EXAMPLE 11 High or low pCO2
ü
“Delayed”
category problems cause PATIENT injury only after many minutes to hours have passed.
EXAMPLE 12 Failure of an infusion pump for
maintenance of intravenous fluids
EXAMPLE 13 Failure of an enteral feeding pump
EXAMPLE 14 Failure of a PATIENT weighing system
The choice of priority should be based upon RISK ANALYSIS.
?
In general,
the lowest priority compatible with the RISK ANALYSIS should be
selected.
ü
In
particular, HIGH
PRIORITY ALARM SIGNALS should be
reserved for those few ALARM
CONDITIONS that truly require immediate response for PATIENT safety—that
is, a response within seconds to a couple of minutes. Many types of equipment
will not require any HIGH
PRIORITY ALARM SIGNALS.
ü
ME EQUIPMENT ALARM SYSTEMS are a
protective measure used to minimize risks to PATIENT, personnel, and
equipment. In certain therapeutic ME
EQUIPMENT, a HAZARDOUS SITUATION could
develop so rapidly, and cause injury or damage so rapidly, that OPERATOR response to even a
well-designed ALARM SYSTEM would be too slow. In such ME
EQUIPMENT, an automatic system of
mitigating the HAZARDOUS SITUATION is highly desirable, if not essential. The general standard and
many particular standards require such safety mechanisms.
ü
It is recognized, however, that no ME EQUIPMENT could have protection
against every possible HAZARD, or in the presence of multiple fault conditions.
יחד עם
זאת, תכן
המערכת צריך
להתחשב
במצבים חריגים
שהם לכאורה
בלתי צפויים