Resetting the alarm after the alarm condition is over
The usability failure mode |
Recommended protection |
Validation |
Sound disabled after resumptionThe user might forget to re-enable the sound after the alarm condition is over IEC 60601-1-8 (section 6.8.2) suggests that a reminder signal may be used, and warns from undesired changes in the settings of these signals. |
Visual indication of disabled soundThe system should provide a visual indication that the sound alarm is disabled. |
Usability
test in training mode:
|
Automatic sound enablingFollowing a delay (of about 30 seconds?) after the alarm condition is over, the sound alarm should be automatically re-enabled. A test sound may be used to notify about the termination of the alarm condition, and the normal monitoring resumption. |
Functional test in training mode |