Muting the sound alarms, worrying about the need to re-enable it
The usability failure mode |
Recommended protection |
Validation |
DistractionThe sound alarm might disturb the treatment |
MutingIn order to enable the medical staff concentrate and focus on the patient condition, they should be able to mute the sound alarm. |
Usability testing in training mode |
Improper mutingThe international standard allows having the following functions that result in muting the alarm:
The user might choose the wrong option, resulting in undesired behavior of the alarm system. |
Consistent mutingOnly one feature should be provided for muting, ensuring consistency. To prevent hasty settings of the alarm limits, the setting operation should be easier than silencing the alarm. |
Functional testingBy inspection of the user interface |