What to check about the user's risks in not noticing the alarm signal
The usability failure mode |
Recommended protection |
Validation |
The user does not notice the sound alarmIf the alarm signal is similar to sound generated by other equipment, then the users might fail to notice it. |
Alarm perception assuranceThe system should always provide a special visual alarm, visible all around the patient, indicating that the alarm is muted. |
Usability testingUse the Training mode for the testing Make sure that the sound alarms do not resemble any equipment used in the patients' surrounding, such as microwave oven. |
The user does not notice the visual alarmIEC 60601-1-8 (for example, section 6.3.2) permits the use of either visual or sound alarms. |
Visual backupSalient visual signals should always accompany the sound alarm signals.
|
Usability testingUse the Training mode for the testing Make sure that the visual alarm signals are visible from all around the patient. |