Forgetting to return to a patient after leaving the patient for a while
The usability failure mode |
Recommended protection |
Validation |
Sound remains mutedThe user might forget that the sound is muted while the treatment has not finished. IEC 60601-1-8 (section 6.8.2) suggests that a reminder signal may be used, and warns from undesired changes in the settings of these signals. |
Visual reminderIf the sound alarm is disabled during an alarm condition, the system should provide a special visual alarm, visible all around the patient, indicating that the alarm is muted. |
Usability
test in training mode:
|