What to check about the user's risks about the generation of alarm signals in alarm conditions
The usability failure mode |
Recommended protection |
Validation |
Alerting is unintentionally disabled.The system might not alert because it is in a special mode, and nobody notices it. |
The system should always alert whenever:
|
Functional testing:Use the Training mode for the testing Wait for the alarm condition, then make sure that the system provides a salient visual alarm signal. |
Risk of alarm failureIf either the visual or sound alarm is disconnected, then the users might know about it only after it is too late. |
Initial alarm testingOn connecting the patient to the alarm system, the system should provide a test sound alarm. A visual alert, visible from all around the patient, should accompany the sound alarm, for testing the visual backup of the sound alarm, and for verification in case that the sound alarm fails. |
Functional testing:
|
Case of sound failureThe international standard suggests that the alarm system may provide continuous information signals, such as heart pulse waveform. Apparently, the standard concern is about situation of sound failure. The risks are about the vigilance of the medical staff. |
Sound detectorIf the alarm system is equipped with means for detecting sound failure, then information signals should be optional.
|
Functional testing
|
Too weak soundThe sound might be too weak compared to the background noise, and distant users might fail to notice it. |
Audibility assuranceThe system should analyze the generated sound compared to the background noise and automatically adjust the sound level to an about-optimal signal-to-noise ratio. A special test control, such as a button, may be used, to always test if the sounds are audible. Alarm-Reliability.pdf |
Functional testing
|