Adjusting the alarm limits to the particular situation of a particular patient. After zooming in to view the sensor values, the users may need to change the alarm limits, to either increase or decrease sensitivity.
The usability failure mode |
Recommended protection |
Validation |
Setting improper alarm limitsThe user might set wrong limits, resulting in either unintentional muting the alarms or in enabling excessive nuisance alarms |
Adjustment by risk levels
|
Usability testingCheck that the changing of alarm limits is easier than silencing the system. |
Failure to control the systemThe international standard suggests (section 6.7), that access to changes to the alarm system may be restricted. The risks of excessive restrictions is that the users might fail to control the system. |
Preventing unintentional mutingOnly risky user actions should be protected. Only administration-level actions should be protected by password. Patient-level actions may be protected by warnings.
|
Functional testingMake sure that the system does not propose risky changes in the alarm limits |