A.1
General guidance
A.1.1 Overview
This annex provides a
rationale for the important requirements of this collateral standard. Its purpose
is to promote effective application of the standard by explaining the reasons
for the requirements, providing examples of how they address certain
alarm-related HAZARDS and providing additional guidance where appropriate.
From the standpoint of PATIENT safety, ALARM SYSTEMS can be hazardous for PATIENTS or OPERATORS if they fail to
effectively warn of potential or actual HAZARDS, cause inappropriate responses, reduce vigilance or interfere
with the performance of the OPERATOR, RESPONSIBLE ORGANIZATION, or other persons.
In addition, in this annex text in italics
indicates guidance that describes means to achieve the safety objectives of
this collateral standard.
A.1.2 ALARM SYSTEMS
As part of the RISK MANAGEMENT PROCESS, the MANUFACTURER identifies RISK CONTROL measure(s) that are
appropriate for reducing the RISK(S) to an acceptable level.
Risk control consists of
an integrated approach in which the MANUFACTURER
uses one or more of the following in the
priority order listed.
a) inherent safety by design;
b) protective measures in the equipment;
c) information for safety, e.g., warnings
and instructions for use, values of monitored variables.
ALARM SYSTEMS as described in this
collateral standard, address b) and c) above by communicating information that
requires a response or awareness by the OPERATOR. The following general principles apply.
d) The ALARM SYSTEM should result in a
greater probability that the OPERATOR will correctly detect and appropriately respond to the condition
that requires their awareness or action than would be the case in the absence
of the ALARM SIGNALS. NOTE Causing
too many ALARM SIGNALS from FALSE POSITIVE ALARM
CONDITIONS can reduce the
effectiveness of an ALARM SYSTEM.
e) ALARM SIGNALS should indicate the
onset and continuing presence of an ALARM
CONDITION.
f) ALARM CONDITIONS should be prioritized
based on the urgency of the required OPERATOR
response (or awareness).
g) Alarm signals should
help the operator:
– determine the urgency of the response required;
– locate the room or part of the room where a response or
awareness is required;
– locate the specific PATIENT or equipment where a response or awareness is required;
– determine or categorize the cause of the ALARM CONDITION; and
– determine or categorize the nature of the response or awareness
that is required.
h) The algorithms that
determine ALARM CONDITIONS should be designed to minimize the number of FALSE NEGATIVE and FALSE POSITIVE ALARM CONDITIONS. Both FALSE NEGATIVE and FALSE POSITIVE ALARM
CONDITIONS are potentially
hazardous. Too many true positive but unhelpful ALARM SIGNALS can result in
inappropriate OPERATOR action or reduce vigilance. Algorithms that determine ALARM CONDITIONS should
be carefully optimized to provide, on balance, an overall benefit to PATIENT care.
i) ALARM SYSTEMS that are continuously attended by an OPERATOR in NORMAL USE should have different
characteristics from ALARM SYSTEMS that are unattended by the OPERATOR
in NORMAL
USE.
j) The design of an ALARM SYSTEM should be based on the TRAINING
and skill of the OPERATOR who is intended to use it.
k) The ALARM SYSTEM should reflect the
problems and needs of the intended environment of use.
l) ALARM SIGNALS should not be
excessively intrusive or degrade the performance of the OPERATOR.
A.1.3 Algorithm quality and performance
ALARM SYSTEM
algorithms should aim at approaching 100 %
sensitivity and 100 % specificity. [7],[8],[9],[10] The leading reason for
disabling ALARM SIGNALS is the
large number of ALARM
SIGNALS associated with FALSE POSITIVE ALARM CONDITIONS, unhelpful
ALARM CONDITIONS, or nuisance ALARM
CONDITIONS. Nuisance ALARM
CONDITIONS are true positives that are unhelpful because
they indicate states that the OPERATOR is already aware of or does not need to know about. [11]
They commonly occur when the ALARM
LIMITS have been set inappropriately
close to an acceptable value but also occur when multiple redundant ALARM CONDITIONS occur in
response to a single underlying problem. Often, ALARM SIGNALS are more confusing than enlightening. Many OPERATORS respond to ALARM SIGNALS by disabling
the ALARM SYSTEM or by
adjusting an ALARM LIMIT to such an
extreme value that the ALARM
SYSTEM is effectively disabled. [12]
Where practical, MANUFACTURERS or writers
of particular standards are encouraged to utilize standardized
physiological databases to validate the algorithms used to determine ALARM CONDITIONS.
Determining and reporting the FALSE
POSITIVE and FALSE
NEGATIVE ALARM CONDITION accuracy in
a standardized format allows OPERATORS and RESPONSIBLE
ORGANIZATIONS to
understand the performance of equipment.
EXAMPLE ANSI/AAMI
EC57:1998, Testing and reporting performance results of
cardiac rhythm and ST segment measurement
algorithms.[5]
Other techniques to reduce the number of FALSE POSITIVE and FALSE NEGATIVE ALARM CONDITIONS include:
a) marking the ALARM SYSTEM with symbol
ISO 7000-0435 when an algorithm cannot classify or
resolve the available data; or
b) using an
ALARM CONDITION DELAY to delay the generation ALARM
SIGNALS for an ALARM CONDITION
to ensure that it remains valid.