6.8.3 * Global indefinite ALARM SIGNAL inactivation states - Rational

ü        If deemed acceptable by RISK ASSESSMENT with regard to the intended environment of use of the ALARM SYSTEM, a global ALARM OFF or AUDIO OFF may be provided. If an ALARM SYSTEM is provided with a global ALARM OFF or AUDIO OFF, the ALARM SYSTEM shall be provided with:

a) a REMINDER SIGNAL; and

b) means to configure (enable or disable) any global ALARM OFF or AUDIO OFF. Such means shall be restricted to the RESPONSIBLE ORGANIZATION and shall prevent the clinical OPERATOR from changing the configuration in NORMAL USE (see 6.7).

וגם את מרווח הזמנים עד לתזכורת

NOTE 1 A global ALARM OFF or AUDIO OFF ALARM SIGNAL inactivation state affects all PHYSIOLOGICAL ALARM CONDITIONS in an ALARM SYSTEM with multiple PHYSIOLOGICAL ALARM CONDITIONS.

זה צריך להיות בפרק ההגדרות

NOTE 2 See also 6.8.2 for requirements for REMINDER SIGNALS.

û       Compliance is checked by inspection.